Due to modified browser settings (e.g. for cookies, JavaScript or browser history), this page cannot be used to its full extent. Please check your settings in order to use the M&C website in the best possible way.
The M&C liquid particle filters CLF-5, CLF-5/W are suitable for the filtration of fluid particles of all types and are recommended for sample gases with an acid dew point above 100 °C. Examples are measurements in flue gas of heavy oil and black coal combustions. The filter separates the aerosols (very fine fluid particles) which still pass the gas cooler. The most effective position of the CLF-... filter is downstream the sample conditioning, directly in front of the flow meter of the analyzer. For additional system protection, we provide the version CLF-5/W with integrated hydrophobic diaphragm, working as a liquid stop.
Liquid particle filter type CLF‑5/W for the separation of microscopically small liquid gas droplets with water stop, glass‑fiber filter element with a retention rate of 99.9999 % for particles > 0.1 µm, flow rate: max. 200 Nl/h, pressure: max. 2 bar abs., sample gas temperature: max. 80 °C, connections: sample gas IN/OUT: G 1/4" f, condensate OUT: GL 25, material: PVDF, glass, Viton®