Heated Sample Line 5N6/8-W
Part No.: 03B5030W
Product features
Electrically heated sample line completely assembled according to your specifications
Inner tube is non-replaceable
Operating temperature: max. +100 °C [+212 °F]
Available up to a length of 30 m (sold by the meter)
Inner tube made of stainless steel with Ø DN 6/8
Outer jacket is a black corrugated PA hose
Operating pressure: max. 100 bar g
Power: 230 V supplied by external temperature controller, 115 V (up to a length of 13 m) available as option
Optionally to a length of 23 m: connection fitting type E with 1.5 m electrical cable and 7 pin-plug for max. 10 A, temperature controller TRD-H1, end fitting type W
Optionally for a length between 23.1 and 30 m: connection fitting type E with 1.5 m electrical cable and 5 pin-plug for max. 20 A, temperature controller TRD-H3, end fitting type W
The electrically heated, plug-in M&C sample lines series 5-N/M/H are used in gas analysis systems to avoid a distortion of measuring values and they also serve as anti-freeze protection during the transport of sample gas from the sample point to the gas conditioning system or directly to the heated analyzer (above dew point).